Our White Paper on Nerve Cuffs represents a basic guideline for working with cuff electrodes in the peripheral nervous system.
It gives an overview about
Leave your name and email address and download the White Paper for free. You will be redirected to the download page.
Our team supports every step from your idea to a medical device. Learn more about how we support you on your journey.
Read more about the experiences of our customers with our products and services in our partnership section.
We are excited to join you on your medical device development journey. Get in contact with our team.
We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.
Please fill out the form below and our sales team will be in contact with you shortly.
FEATURE
Recording channels
Sampling rate
Sampling dynamic range
High pass filter cut-off
Low pass filter cut-off
Amplifier band pass gain
Band pass roll-off
Reference
Stimulation
Stimulation channels
Current
Current source
Pulse width
Power supply
Wireless data transmission
Closed Loop latency
VALUE
32
1 kHz
16 bit (74 nV smallest increment)
ca. 2 Hz
325 Hz
Adjustable: 100-750
20 dB/dec
Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact
Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width)
32
Max. -6 mA / +1.5 mA (24 µA increments) within
compliance voltage range of -11 V to +5 V
Can be directed to any of the 32 electrode contacts
Negative phase: 10 µs – 2,500 µs
Wireless inductive, 120-140 kHz
Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms