High Channel Packaging

Ceramic Implant Housing

Our advanced encapsulation solutions provide hermetic sealing and robust insulation, enhancing the durability and performance of implantable electronics.

Miniature Designs

With custom made minitaure designs we enable smaller, lighter, and less invasive implant designs, improving patient comfort and facilitating advanced therapeutic applications.

Reduced size also means lower power consumption, enhanced integration with other components, and greater design flexibility for next-generation medical devices.

High Channel Encapsulation

Based on our Brain Interchange System we offer high-channel hermetic encapsulations. Ceramic cover can be replaced by metal cover if needed.

These are optimized solutions for complex signal processing, allowing you to increase precision, resolution, and functionality in your specific application.

Beyond encapsulation, our expertise extends to silicone overmolding, offering flexible, biocompatible protection, as well as custom cable and electrode solutions for seamless connectivity in complex implant systems.

Explore our full range of technologies designed to support the next generation of active implantables.

Custom Design

Thank you for your interest in a custom design from CorTec.

Please fill out the form below and our team from sales and engineering will be in contact with you shortly.

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms