Percutaneous Electrodes

The °AirRay Percutaneous Electrodes are designed to provide an electrical interface to subcutaneous, intraspinal or deep brain locations of the nervous system. All contacts are able to record and stimulate.

Arrangement of 360° electrode contacts on flexible or pushable tubes for subcutaneous or spinal cord recording and stimulation.

We offer various designs for your specific application:

  • Number of Contacts
  • Length and thickness of the silicone tube
  • Contact placement
  • Thickness, Stiffness, Softness

°AirRay Percutaneous Electrode | 4 Contact Rings 360°

  • 4 Contact Rings
  • Contact Length: 3 mm
  • Contact Diameter: 1 mm

Suture Sleeves

Designed to anchor the implanted electrode wires within surrounding tissue. Key benefits of our Suture Sleeves include:

  • Secure Ligature: Prevents electrode movement and displacement
  • Radio-opaque: Clearly visible in all kinds of X-rays
  • Biocompatible: Safe for use in biological environments

Custom Design

Thank you for your interest in a custom design from CorTec.

Please fill out the form below and our team from sales and engineering will be in contact with you shortly.

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms