Next Important Milestone Reached for First-in-Human Implantation of the CorTec Brain Interchange System as a Therapeutic Brain-Computer Interface

CorTec has announced today that Institutional Review Board from University of Washington (IRB) has approved the Feasibility study for stroke rehabilitation using the Brain Interchange System. This marks a significant milestone and paves the way for the system’s clinical investigation in the United States. The IRB approval follows the recent Investigational Device Exemption (IDE) granted by the U.S. Food and Drug Administration (FDA), setting the stage for the study to proceed to participant recruitment.

IRB approval is a critical step in the clinical research process, especially for studies involving human subjects. While FDA approval ensures the safety and technical readiness of the device being studied, the IRB provides an essential layer of oversight to safeguard the ethical conduct of the research. The IRB’s review takes into account the specific local context of the research site, including the population being studied and the resources available, to ensure participant safety. This local oversight is crucial, as it tailors the ethical review process to the unique circumstances of each study, beyond the federal standards.

“Securing FDA approval for our IDE study(1) was a thorough process that involved the submission of extensive documentation from both our team at the University of Washington and the device manufacturer, CorTec,” said Dr. Jeffrey Herron, Assistant Professor at the University of Washington and co-investigator on the NIH-funded study(2). “The FDA’s evaluation focused on the study protocol, hazard analysis, and the device design to ensure readiness for human implantation. Achieving this IDE approval was a critical milestone, and with the following IRB approval, we are now prepared to move forward with participant recruitment.”

According to Dr. Martin Schuettler, CorTec’s CTO, the Brain Interchange System‘s closed-loop functionality is a groundbreaking achievement that opens new avenues for highly individualized treatments. “I am incredibly proud of this milestone,” Dr. Schuettler expressed. “Our system enables a seamless exchange of information between biology and technology, which is why we named it the CorTec Brain Interchange. With this system, we are not only providing the critical technological tools needed to advance new therapies but also shaping the future of brain-computer interfaces and paving the way for future therapy discoveries.”

CorTec is collaborating with leading experts in the U.S. to further the development of innovative therapies using the Brain Interchange System. The first Investigational Device Exemption (IDE) study will be conducted in collaboration with Professor Jeffrey G. Ojemann from the University of Washington School of Medicine in Seattle and Professor Steven C. Cramer from the University of California, Los Angeles. Funded by the U.S. National Institutes of Health (NIH), this consortium aims to gather initial safety data and develop novel therapeutic rehabilitation approaches for upper limb impairment in stroke patients through direct cortical electrical stimulation delivered by the Brain Interchange System(2).

Patient enrollment and the first implantation of the neural interfacing system are scheduled for the fourth quarter of 2024, marking the beginning of a new chapter in stroke rehabilitation research.

 

Go to the PDF download of this press release (English Version).

Go to the PDF download of this press release (German Version).

 

Media Contact:

Carolina Remke – Head of Marketing
pr@cortec-neuro.com
www.cortec-neuro.com

______________________________________________________________________________________
Disclaimer: The research reported in this publication is supported by the National Institute Of Neurological Disorders And Stroke of the National Institutes of Health under Award Number UH3NS121565. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.
______________________________________________________________________________________
References:
[1] IDE Application reference G230003/A001
[2] NIH Project 1UH3NS121565-01A1: ” Motor Recovery through Plasticity-Inducing Cortical Stimulation”, in response to RFA-NS-18-023: https://reporter.nih.gov/search/-xvTvG85Ukm-KXyunAWaJw/project-details/10357993

 

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms