Live podcast at the Käpsele Innovation festival with our CTO

The ‘Käpsele Innovation Festival’ is taking place on the 26th of June at the Sick Arena in Freiburg. Companies from Freiburg, Germany and Europe will showcase their innovative ideas and companies. We will also be there to share our expertise in a live podcast. Our CTO Martin Schuettler will be on stage together with Prof. Dr. Thomas Stieglitz, Managing Director at IMTEK, to demonstrate the possible applications of brain implants.

Come along and draw inspiration from creative minds and ambitious innovations.

You can find more information and tickets here: https://www.innovation-festival.de/

You might also be interested in the following events:

June, 15-18

HLTH Europe 2026

Amsterdam, Netherlands
June 11, 2026

SmartXhealth

Freiburg, Germany
April 28-30, 2026

Innovation Summit Dublin 2026

Dublin, Irland
Mar 25-26, 2026

LSX World Congress Europe

Lisbon, Portugal
Mar 16-20, 2026

LSI USA 2026

Dana Point, USA
March 3-5, 2026

health.tech Global Summit in Basel

Basel, Switzerland

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms