CorTec °AirRay Cuff Electrodes – Whitepaper

Some time ago we published a Whitepaper for the use of our °AirRay Cuff Electrodes. Not only gives it useful insights about choosing the right Cuff for your specific application but also provides a lot of intel about the tricks in using them.

From preparation to explantation – here you will find anything you need to know.
Download the free Whitepaper here.

Further readings in Technology:

The Power of Active Implant Sytems

Active implant systems are the foundation of modern neurotechnology. It enables precise, bidirectional communication with...

Interview with Mariska van Steensel: Inside the INTENSE Project

Dr. Mariska van Steensel from University Medical Center Utrecht shares insights from the INTENSE Project,...

CorTec: Your Partner for Fully Integrated Neuromodulation & Medical Device Solutions 

Bringing a neuromodulation medical device to clinical reality is rarely just engineering. It is a...

Brain Interchange System Sets a New Standard

The recent publication in Nature Scientific Data by the research group around Takufumi Yanagisawa at...

Contact

Do you have any questions regarding our Company, our Products or our Technology?
We are happy to receive your message!

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms