renaboehringer

CorTec’s Brain Interchange™ BCI System Enables Stroke Patient to Control Computer with his Mind

The same fully implanted, wireless device supporting motor recovery after stroke now enables thought-controlled computer use without additional surgery or hardware.   Freiburg, Germany, April 29, 2026 – CorTec GmbH today announced that the first participant in the University of Washington’s NIH-funded clinical trial of the Brain Interchange™ BCI system has successfully controlled a computer […]

CorTec’s Brain Interchange™ BCI System Enables Stroke Patient to Control Computer with his Mind Read More »

CorTec’s Brain Interchange Accepted into FDA Total Product Life Cycle Advisory Program (TAP) – Third patient successfully implanted at Harborview Medical Center as NIH-funded stroke trial advances

Freiburg, Germany, April 23, 2026 – CorTec GmbH today announced that the U.S. Food and Drug Administration (FDA) has accepted the company into its Total Product Life Cycle Advisory Program (TAP), recognizing the Brain Interchange™ BCI system as a priority innovation in stroke motor rehabilitation. The TAP enrollment follows the recently granted FDA Breakthrough Device

CorTec’s Brain Interchange Accepted into FDA Total Product Life Cycle Advisory Program (TAP) – Third patient successfully implanted at Harborview Medical Center as NIH-funded stroke trial advances Read More »

CorTec Announces Successful Second Human Implantation of Its Brain-Computer Interface (BCI) System

Second implantation at Harborview Medical Center in Seattle marks continued progress in the FDA-approved study informed by promising results from the first participant. Freiburg, Germany, February 10, 2026 – CorTec GmbH, a pioneer in active implantable medical technologies, today announced the successful second implantation of its proprietary BrainComputer Interface (BCI) system, the Brain Interchange™, in

CorTec Announces Successful Second Human Implantation of Its Brain-Computer Interface (BCI) System Read More »

2025 in Review

Looking back, 2025 has been a truly remarkable year for CorTec, filled with milestones, inspiring collaborations, and unforgettable moments.   Early in the year, we entered into a strategic partnership with Heraeus Medevio. Together combine our strengths in development, prototyping, small series production and large series production.  This way both companies create scalable, high-quality solutions,

2025 in Review Read More »

CorTec Announces Neurotech Milestone: First Human Implantation of a Brain-Computer Interface made in Germany

The closed-loop BCI system, developed entirely in Germany, is currently being evaluated in an FDA-approved clinical trial with stroke patients at Harborview Medical Center in Seattle, a part of UW Medicine   Freiburg, Germany, July 29, 2025 – CorTec GmbH, a pioneer in active implantable medical technologies, today announced the first human implantation of its

CorTec Announces Neurotech Milestone: First Human Implantation of a Brain-Computer Interface made in Germany Read More »

Our Brain Interchange BCI System Reaches Clinical Readiness

CorTec’s fully implantable Brain-Computer Interface (BCI) system has reached clinical readiness, and we are in preparation for the next milestone on our journey – the first in human implantation. The system is based on a modular platform that enables flexible use across a range of neurotechnological applications. With high-resolution signal processing and bidirectional communication, the

Our Brain Interchange BCI System Reaches Clinical Readiness Read More »

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms