Our °AirRay Electrode Technology allows us to manufacture electrodes that are tailored to your specific needs. The degree of flexibility of the electrodes plays an important role for many applications.

Our proprietary laser-aided manufacturing process enables us to vary the flexibility of our electrodes from very soft to harder versions.

Our new video gives you an impression about the behavior of electrodes with different degrees in flexibility.

Watch the video now and explore the possibilities for your own application.

Further readings in Technology:

Interview with Mariska van Steensel: Inside the INTENSE Project Part 2

Dr. Mariska van Steensel from University Medical Center Utrecht shares further insights from the INTENSE...

From Leads to System Integration

The lead is the starting point of every active implant system. Its geometry, materials, and...

The Power of Active Implant Sytems

Active implant systems are the foundation of modern neurotechnology. It enables precise, bidirectional communication with...

Interview with Mariska van Steensel: Inside the INTENSE Project

Dr. Mariska van Steensel from University Medical Center Utrecht shares insights from the INTENSE Project,...

Contact

Do you have any questions regarding our Company, our Products or our Technology?
We are happy to receive your message!

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms