These days the Joint Congress of European Neurology takes place in Istanbul. In the European Year of the Brain which the European Brain Council (EBC) has procalimed for 2014, neurologists emphasise an increase of neurological diseases. According to EBC surveys, over 200 milion people in Europe with increasing tendency are suffering from partly serious neurological

Congress of European Neurology: Increase of neurological diseases Read More »

Zunahme neurologischer Erkrankungen fordert verstärkte Aktivitäten

Dieser Tage findet in Istanbul der Joint Congress of European Neurology statt. Im Europäischen Jahr des Gehirns, das der European Brain Council (EBC) für 2014 ausgerufen hat, weisen Neurologen auf die Zunahme neurologischer Erkrankungen hin. Über 200 Millionen Menschen in Europa mit steigender Tendenz leiden nach Erhebungen des EBC an teils schwerwiegenden neurologischen Erkrankungen, die

Zunahme neurologischer Erkrankungen fordert verstärkte Aktivitäten Read More »

Get a Quote

We’re happy to help you realize your next project. Whether it is a complete implantable system design, an electrode or another component.

Please fill out the form below and our sales team will be in contact with you shortly.

SPECIFICATIONS

FEATURE 

Recording channels 

Sampling rate 

Sampling dynamic range 

High pass filter cut-off 

Low pass filter cut-off 

Amplifier band pass gain 

Band pass roll-off 

Reference


Stimulation 

Stimulation channels 

Current 

Current source 

Pulse width 

Power supply 

Wireless data transmission 

Closed Loop latency

VALUE

32 

1 kHz 

16 bit (74 nV smallest increment) 

ca. 2 Hz 

325 Hz 

Adjustable: 100-750 

20 dB/dec 

Any (subset) of the recording channels selectable by software or one dedicated hard-wired additional contact 

Current-controlled, biphasic, rectangular, asymmetric stimulus pulses (cathodic amplitude with pulse width followed by an anodic counter pulse of 1/4x amplitude and 4x pulse width) 

 32 

Max. -6 mA / +1.5 mA (24 µA increments) within

 compliance voltage range of -11 V to +5 V 

Can be directed to any of the 32 electrode contacts 

Negative phase: 10 µs – 2,500 µs

Wireless inductive, 120-140 kHz

Bi-directional, radio frequency in 2400-2483.5 MHz band ≤ 40 ms